Beckman Coulter, Inc.: Medical Device Recall in 2025 - (Recall #: Z-2270-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461
Class II
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
UDI-DI: 15099590369248; Serial Numbers: 902101
Worldwide distribution - US Nationwide.
Voluntary: Firm initiated