Becton Dickinson & Co.: Medical Device Recall in 2015 - (Recall #: Z-0974-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar Trypticase Soy Agar with 5% Sheep Blood is used for the growth of fastidious organisms and for the visualization of hemolytic reactions. MacConkey II Agar is a selective and differential medium for the detection of coliform organisms and enteric pathogens.

Product Classification:

Class II

Date Initiated: June 6, 2014
Date Posted: January 21, 2015
Recall Number: Z-0974-2015
Event ID: 69091
Reason for Recall:

A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as 1-3 CFUs per plate when present.

Status: Terminated
Product Quantity: 7,617 cases (100 plates/case)
Code Information:

Catalogue # 221291 with Lot #4050354, 4086399, 4099206 and 4143653; Catalogue #221261 with Lot # 4022158 4086443 4092199

Distribution Pattern:

Worldwide Distribution - US nationwide including Puerto Rico, Belgium, Canada, Philippines, Singapore, and Taiwan.

Voluntary or Mandated:

Voluntary: Firm initiated