Becton Dickinson & Co.: Medical Device Recall in 2015 - (Recall #: Z-0976-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Columbia Agar with 5% Sheep Blood Columbia Agar with 5% Sheep Blood is a general purpose culture medium used for the growth and recovery of fastidious microbial species.
Product Classification:
Class II
Date Initiated: June 6, 2014
Date Posted: January 21, 2015
Recall Number: Z-0976-2015
Event ID: 69091
Reason for Recall:
A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as 1-3 CFUs per plate when present.
Status: Terminated
Product Quantity: 775 cases
Code Information:
Catalogue # 221263 with Lot #4099228
Distribution Pattern:
Worldwide Distribution - US nationwide including Puerto Rico, Belgium, Canada, Philippines, Singapore, and Taiwan.
Voluntary or Mandated:
Voluntary: Firm initiated