Becton Dickinson & Co.: Medical Device Recall in 2021 - (Recall #: Z-1377-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221558

Product Classification:

Class II

Date Initiated: March 17, 2021
Date Posted: April 14, 2021
Recall Number: Z-1377-2021
Event ID: 87546
Reason for Recall:

Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment

Status: Terminated
Product Quantity: 15550 EA
Code Information:

Lot Numbers: 0317559 Exp. Date 2021-MAR-11; 0345695 Exp. Date 2021-APR-06 UDI: 01)30382902215586 (10)0317559(17)031121 (30)0100 (01)30382902215586 (10)0345695(17)040621 (30)0100

Distribution Pattern:

Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.

Voluntary or Mandated:

Voluntary: Firm initiated