Becton Dickinson & Co.: Medical Device Recall in 2022 - (Recall #: Z-0232-2023)
See the recall detail below. You can also see other recalls from the same firm in 2022.
BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) Number: 448607
Class II
Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic
UDI: (10) 1355046 (17) 230131 (30) 0025 (01) 30382904486076 Lot Number: 1355046 Exp. Date: 31JAN2023
US Nationwide distribution.
Voluntary: Firm initiated