Becton Dickinson & Co.: Medical Device Recall in 2024 - (Recall #: Z-1029-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
BD BBL Sensi Disc Doripenem- 15 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 232219¿
Class II
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Catalog No. 232219¿ UDI-DI 30382902322192 Lots 3040749 3093444 3163515 3296612
Domestic distribution nationwide. International distribution worldwide.
Voluntary: Firm initiated