Becton Dickinson & Co.: Medical Device Recall in 2024 - (Recall #: Z-1031-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

BD BBL Sensi Disc Imipenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231644 and 231645¿

Product Classification:

Class II

Date Initiated: January 8, 2024
Date Posted: February 14, 2024
Recall Number: Z-1031-2024
Event ID: 93789
Reason for Recall:

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Status: Ongoing
Product Quantity: 2,363,168 total units
Code Information:

Catalog No. 231644 UDI-DI 00382902316442 Lots 1335006 2056580 2091918 2210318 2301385 3062101 3094134 3118125 3214950; Catalog No. 231645¿ UDI-DI 30382902316450 Lots 3062101 3094134 3118125 3214950 3261103 1335006 2056580 2081151 2091918 2091925 2147704 2210318 2210343 2277500 2301385

Distribution Pattern:

Domestic distribution nationwide. International distribution worldwide.

Voluntary or Mandated:

Voluntary: Firm initiated