Becton Dickinson & Company: Medical Device Recall in 2014 - (Recall #: Z-0772-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

BD Interlink threaded lock cannula This is a locking blunt plastic cannula specifically designed for IV connection to an Interlink¿ injection site. It is screwed down over the Interlink¿ injection site and a blunt plastic cannula inside the housing enters the Interlink¿ injection site to provide a pathway for IV fluid administration.

Product Classification:

Class II

Date Initiated: November 26, 2013
Date Posted: January 22, 2014
Recall Number: Z-0772-2014
Event ID: 67098
Reason for Recall:

BD Interlink Threaded Lock Cannula REF# 303369 and lot #3021392 is being recalled due to the potential that a small percentage of the lot may exhibit open seals which adversely impacts unit package integrity and product sterility.

Status: Terminated
Product Quantity: 268,500 products
Code Information:

Marketing Status: Class II, 510K # K011858 Device Listing Number: D043644 Lot number 3021392

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Canada and Japan.

Voluntary or Mandated:

Voluntary: Firm initiated