Becton Dickinson & Company: Medical Device Recall in 2016 - (Recall #: Z-0571-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 4.0 mL (Lavender) BD Hemogard Additive: K2EDTA (spray dried), 7.2mg Sterile IVD BD, Made in USA BD Vacutainer¿ tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory. The BD Vacutainer¿ Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening.
Class II
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
Catalog Reference #367844; Lot number 5154624 Catalog Reference #367862; Lot numbers 5089771 and 5187753 Catalog Reference #368021; Lot number 5089820
Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
Voluntary: Firm initiated