Becton Dickinson & Company: Medical Device Recall in 2017 - (Recall #: Z-1746-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.

Product Classification:

Class II

Date Initiated: March 16, 2017
Date Posted: April 12, 2017
Recall Number: Z-1746-2017
Event ID: 76826
Reason for Recall:

BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.

Status: Terminated
Product Quantity: 13,064,500 tubes (non-expired)
Code Information:

Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date.

Distribution Pattern:

Worldwide - US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated