Becton Dickinson & Company: Medical Device Recall in 2017 - (Recall #: Z-2750-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

BD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #309580 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection

Product Classification:

Class II

Date Initiated: September 16, 2016
Date Posted: July 26, 2017
Recall Number: Z-2750-2017
Event ID: 77513
Reason for Recall:

BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.

Status: Terminated
Product Quantity: 340,000 units
Code Information:

6090832 2021-03 6176715 2021-06

Distribution Pattern:

Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated