Becton Dickinson & Company: Medical Device Recall in 2022 - (Recall #: Z-1437-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

BD Connecta" Plus1 360 Blue Blend Catalog No. 394911 (OUS)

Product Classification:

Class II

Date Initiated: May 9, 2022
Date Posted: July 27, 2022
Recall Number: Z-1437-2022
Event ID: 90304
Reason for Recall:

May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.

Status: Ongoing
Product Quantity: 9,010
Code Information:

GTIN: 382903949113 Lot Numbers/Exp.Date: 1061512 2/29/2024

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of AK, AR, AZ, CA, CO, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WV and the countries of Belgium, China, India, Indonesia, Japan, Korea, Nepal, Philippines, Singapore, Sri Lanka, Thailand, Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated