Belmont Instrument LLC: Medical Device Recall in 2025 - (Recall #: Z-1099-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or alternate care environments to provide warmed blood and fluids to any patients e10 kg requiring warmed infusion. Part Number:903-00006
Class II
Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set
UDI-DI: 00896128002022 Lot Numbers: 20240910, 20241002, 20241006, 20241014
Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Japan, Malaysia, Poland, Switzerland, Taiwan (Province of China), Ukraine.
Voluntary: Firm initiated