Berchtold Corp.: Medical Device Recall in 2016 - (Recall #: Z-1925-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Surgical Table Cushion (HybriGel Pads) HybrGel Pads are an optional accessory to the Operon Surgical Table. The pads may be attached to the base section with a Velcro fastener. The HybriGel pad shapes and variants can differ depending on the scope of the delivery.

Product Classification:

Class II

Date Initiated: April 19, 2016
Date Posted: June 15, 2016
Recall Number: Z-1925-2016
Event ID: 73999
Reason for Recall:

The Surgical Table Cushion (HybriGel Pads) could potentially contain staples that's secure two underlying fabrics, which would be captured on certain scans on patients, such as x-ray.

Status: Terminated
Product Quantity: 227 units
Code Information:

141364, 141365, 141367, 141369, 142575, 142577, 142579, 142769, 142771, 142773, 142774, 142775, 143420, 144088, 144384, 144447, 145384, 145529, 145545, 145936, 146141, 146963, 147304, 147305, 147306

Distribution Pattern:

Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated