Bien Air Dental SA: Medical Device Recall in 2015 - (Recall #: Z-2797-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Contra-Angles CA1:1L EVO 15 Catalog numbers: 1600939-001 and 1600938-001 Used for generally dentistry procedures
Product Classification:
Class II
Date Initiated: August 27, 2015
Date Posted: September 30, 2015
Recall Number: Z-2797-2015
Event ID: 72079
Reason for Recall:
Possible loosening of the dental tool, which may lead to potential swallowing of the tool.
Status: Terminated
Product Quantity: 112 (U.S.)
Code Information:
Catalog numbers: 1600939-001 and 1600938-001 Serial numbers concerned: - Ref: 1600939-001: Serial numbers included in the following range: 15B0001 up to 15F0011 - Ref: 1600938-001: Serial numbers included in the following range: 15B0007 up to 15F0027
Distribution Pattern:
Worldwide Distribution - US distribution to one distributor in FL.
Voluntary or Mandated:
Voluntary: Firm initiated