BioFire Diagnostics, LLC: Medical Device Recall in 2021 - (Recall #: Z-2039-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

Product Classification:

Class II

Date Initiated: June 7, 2021
Date Posted: July 14, 2021
Recall Number: Z-2039-2021
Event ID: 88117
Reason for Recall:

Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.

Status: Terminated
Product Quantity: 4,462 kits (Globally)
Code Information:

Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

Distribution Pattern:

U.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece , Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, Nicaragua, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated