BioMerieux SA: Medical Device Recall in 2017 - (Recall #: Z-0265-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

ETEST OXACILLIN OX 256 WW F100 In vitro diagnostic, a quantitative technique for determining the antimicrobial susceptibility of bacteria species.

Product Classification:

Class II

Date Initiated: April 13, 2017
Date Posted: December 27, 2017
Recall Number: Z-0265-2018
Event ID: 78688
Reason for Recall:

QC failures (MIC out of range high) for S.aureus ATCC 29213 strain on ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) reported. The product behavior could lead to False Resistant results.

Status: Terminated
Product Quantity: 1585 units
Code Information:

REF: 520518, Lot numbers: 1003055340, 1003315740, 1004070580, 1004319590, 1004818850, 1004890270 and 1005366110.

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and to the countries of : Algeria, Austria, Australia, Belarus, Bosnia-Herz., Canada, Croatia, Czech Republic, Egypt, Estonia, France, Georgia, Germany, Greece, Guam, Hong Kong, Hungary, India, Israel, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Nigeria, Peru, Philippines, Qatar, Russian Fed., Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Utd.Arab Emir.

Voluntary or Mandated:

Voluntary: Firm initiated