BioMerieux SA: Medical Device Recall in 2018 - (Recall #: Z-2164-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

The VITEK 2 Gram-Positive identification card (GP) is intended for use with the VITEK 2 Systems for the automated identification of most significant Gram-positive organisms. The VITEK 2 GP identification card Is a single-use disposable. GP TEST KIT VTK2 20 CARTES A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms.

Product Classification:

Class II

Date Initiated: March 23, 2018
Date Posted: June 20, 2018
Recall Number: Z-2164-2018
Event ID: 80111
Reason for Recall:

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Status: Terminated
Product Quantity: 15 boxes with GP TEST KIT VTK2 20 CARDS
Code Information:

VITEK¿ 2 GP (ref: 21342):Lot Number 2420352403

Distribution Pattern:

International distribution.

Voluntary or Mandated:

Voluntary: Firm initiated