Biomet 3i, LLC: Medical Device Recall in 2012 - (Recall #: Z-2158-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

***REF IFNT611***Full OSSEOTITE Tapered Certain Implant 6 x 11.5mm***Sterile by radiation***Rx Only***BIOMET 3i Dental Iberica S.L. WTC Almeda Park, Ed, 1, Planta 1 Pl. de la Pau s/n08940 Cornella de Llobregat (Barcelona) Spain. Lot #2011110798. Intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully endentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridge work and to retain over dentures.

Product Classification:

Class II

Date Initiated: March 23, 2012
Date Posted: August 15, 2012
Recall Number: Z-2158-2012
Event ID: 62375
Reason for Recall:

Biomet 3i is recalling their product Full Osseotite Tapered Certain Implant IFNT611. The depth of the implant's internal hex is too shallow. The condition, if present, may prevent the driver; cover screw or abutment from fully engaging the implant.

Status: Terminated
Product Quantity: 240
Code Information:

Lot Number 2011110798

Distribution Pattern:

Worldwide Distribution -- United States, UK, France, Canada, Mexico, Poland, Lebanon and Hong Kong.

Voluntary or Mandated:

Voluntary: Firm initiated