Biomet, Inc.: Medical Device Recall in 2012 - (Recall #: Z-2486-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Optilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.

Product Classification:

Class II

Date Initiated: August 23, 2012
Date Posted: October 10, 2012
Recall Number: Z-2486-2012
Event ID: 63073
Reason for Recall:

Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw

Status: Terminated
Product Quantity: 363 units
Code Information:

Catalog number: 36505; and lot numbers: 069540, 179190, 201610, 385380, 419940, 427530, 440360, 440370, 555370, 638560, 645820, and 956960.

Distribution Pattern:

Worldwide Distribution-USA (nationwide) including the states of AL, GA, MT, TA, AZ, WA, CA, TA, IN, NY, NJ, VA, MA, MI, MD, PA, KY, NC, UT, NH, FL, OK, CO, LA, OH, MO and SD, and the countries of Japan, Netherlands, Australia, and Costa Rica.

Voluntary or Mandated:

Voluntary: Firm initiated