Biomet, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0450-2016)
See the recall detail below. You can also see other recalls from the same firm in 2015.
A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.
Class II
These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.
Part number and size (non-sterile) 816140010, 4.0X10MM; 816140012, 4.0X12MM; 816140014, 4.0X14MM; 816140016, 4.0X16MM; 816140018, 4.0X18MM; 816140020, 4.0X20MM; 816140022, 4.0X22MM; 816140024, 4.0X24MM; 816140026, 4.0X26MM; 816140028, 4.0X28MM; 816140030, 4.0X30MM; 816140032, 4.0X32MM; 816140034, 4.0X34MM; 816140036, 4.0X36MM; 816140038, 4.0X38MM; 816140040, 4.0X40MM; 816140042, 4.0X42MM; 816140044, 4.0X44MM; 816140046, 4.0X46MM; 816140048, 4.0X48MM; 816140050, 4.0X50MM; 816140055, 4.0X55MM; 816140060, 4.0X60MM; 816140065, 4.0X65MM; 816140070, 4.0X70MM; 816140075, 4.0X75MM; 816140080, 4.0X80MM; 816140085, 4.0X85MM; 816140090, 4.0X90MM; 816140095, 4.0X95MM; 816140100, 4.0X100MM Part number and size (sterile) 856140010, 4.0X10MM; 856140012, 4.0X12MM; 856140014, 4.0X14MM; 856140016, 4.0X16MM; 856140018, 4.0X18MM; 856140020, 4.0X20MM; 856140022, 4.0X22MM; 856140024, 4.0X24MM; 856140026, 4.0X26MM; 856140028, 4.0X28MM; 856140030, 4.0X30MM; 856140032, 4.0X32MM; 856140034, 4.0X34MM; 856140036, 4.0X36MM; 856140038, 4.0X38MM; 856140040, 4.0X40MM; 856140042, 4.0X42MM; 856140044, 4.0X44MM; 856140046, 4.0X46MM; 856140048, 4.0X48MM; 856140050, 4.0X50MM; 856140055, 4.0X55MM; 856140060, 4.0X60MM; 856140065, 4.0X65MM; 856140070, 4.0X70MM. All lots manufactured between 06/01/08 and 06/01/15.
AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR CHILE, COSTA RICA, NETHERLANDS, CANADA
Voluntary: Firm initiated