Biomet, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2153-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Knee Products: 183742 Vanguard Knee System, PS+ Tibial Bearing, 12 MM, 71/75 MM EP-183608 Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM EP-183608 Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM 183744 Vanguard Knee System, PS+ Tibial Bearing, 14 MM, 71/75 MM 183742 Vanguard Knee System, PS+ Tibial Bearing, 12 MM, 71/75 MM EP-183608 Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM EP-183608 Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM Product Usage: Knee prosthesis
Class II
Potential presence of elevated endotoxin levels that exceed the specification limit
Lot Numbers: UDI Number 294130 Knees Vanguard Knee System, PS+ Tibial Bearing, 12 MM, 71/75 MM (01)00880304438262(17)211219(10)294130 509970 Knees Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM (01)00880304476714(17)211230(10)509970 509990 Knees Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM (01)00880304476714(17)220103(10)509990 175980 Knees Vanguard Knee System, PS+ Tibial Bearing, 14 MM, 71/75 MM (01)00880304438354(17)220114(10)175980 294130R Knees Vanguard Knee System, PS+ Tibial Bearing, 12 MM, 71/75 MM (01)00880304438262(17)211219(10)294130R 509970R Knees Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM (01)00880304476714(17)211230(10)509970R 509990R Knees Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM (01)00880304476714(17)220103(10)509990R
Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.
Voluntary: Firm initiated