Biomet, Inc.: Medical Device Recall in 2021 - (Recall #: Z-2261-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Arcos Modular Revision Hip System Standard Cone Prox Body, Size B, 60 MM Item Number: 11-301302
Product Classification:
Class II
Date Initiated: June 2, 2021
Date Posted: August 18, 2021
Recall Number: Z-2261-2021
Event ID: 88304
Reason for Recall:
Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
Status: Terminated
Product Quantity: 21 units
Code Information:
Lot Numbers/UDI : 626540 (01)00880304474123(17)310423(10)626540 679320 (01)00880304474123(17)310419(10)679320
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated