Biomet U.K., Ltd.: Medical Device Recall in 2017 - (Recall #: Z-1157-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Oxford Partial Knee System Left Medial Tibial Trial Tray Size F Product Usage: Instruments for use with the Oxford Uni Partial Knee System

Product Classification:

Class II

Date Initiated: January 5, 2016
Date Posted: March 1, 2017
Recall Number: Z-1157-2017
Event ID: 76179
Reason for Recall:

Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.

Status: Terminated
Product Quantity: 72 units in total
Code Information:

Item: 32-420822, Lots: ZB160501 ZB160601 ZB160701

Distribution Pattern:

Worldwide Distribution in the states of LA, NC, AR, MT, MO, CA, NY, KS, TX, MA, KY, OK, SC, NJ and the countries of Foreign: THAILAND, JAPAN, NETHERLANDS

Voluntary or Mandated:

Voluntary: Firm initiated