Biosense Webster, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1242-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
LASSO NAV Duo Loop eco Catheters, Catalog No. D-1354-01, D-1354-02, D-1354-03, D-1354-04, D-1354-05, D-1354-06. LASSO NAV DUO LOOP eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart. LASSO NAV DUO LOOP eco Catheter provides location information when used with compatible CARTO 3 EP Navigation Systems.
Class II
Biosense Webster is recalling the LASSO NAV Duo Loop eco Catheters because of reports of difficulty retracting and damage to the spine cover.
All lots manufactured as of February 24, 2015 are affected.
Nationwide Distribution
Voluntary: Firm initiated