Boston Scientific Corporation: Medical Device Recall in 2019 - (Recall #: Z-0699-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Product Classification:

Class III

Date Initiated: November 28, 2018
Date Posted: January 16, 2019
Recall Number: Z-0699-2019
Event ID: 81747
Reason for Recall:

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

Status: Ongoing
Product Quantity: 1,282 (all three types in Japan)
Code Information:

Blazer Open Irrigated Ablation Catheter Material Number: M00496200, Batch: 20399570, Expiration Date: 3/13/2020 Batch: 21778776, Expiration Date: 2/21/2021 Batch: 22152470, Expiration Date: 5/21/2021 Batch: 22499259, Expiration Date: 8/13/2001 Material Number: M0049620N40 Batch::20413170, Expiration Date: 3/16/2020 Batch: 20809676, Expiration Date: 6/22/2020 Material Number: M0049620K20 Batch: 20403945, Expiration Date: 3/13/2020

Distribution Pattern:

Japan

Voluntary or Mandated:

Voluntary: Firm initiated