Boston Scientific Corporation: Medical Device Recall in 2020 - (Recall #: Z-1804-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System

Product Classification:

Class II

Date Initiated: September 18, 2019
Date Posted: May 6, 2020
Recall Number: Z-1804-2020
Event ID: 85271
Reason for Recall:

Certain iLab Polaris Systems which were upgraded with a previous 2.10 software version (disk batch 629774) contained a bug in which only the True Color Secondary Capture (SC) format will archive Diastolic Hyperemia-Free Ratio data as intended. If either Multi-Frame UltraSound (US) or SC Image Storage formats are used, the DFR data will not be archived and the user will not be notified of this unexpected behavior.

Status: Completed
Product Quantity: 150 units
Code Information:

Polaris 2.10 software version (disk batch 629774); iLab Polaris Systems affected product listing (serial numbers): 200204, 200209, 200217, 200308, 200310, 200448, 200450, 200456, 200457, 200458, 200461,200480, 200481, 200103, 200145, 200044, 200118, 200360, 200498, 200170, 8361, 10312, 10443, 10548, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278,10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278, 10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278, 10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 6582 , 7023, 7857, 5637, 5716, 6650, 6894, 4031, 5003, 5043, 5060, 5109, 5232, 5239, 5393, 5465, 5651, 5678, 5860, 6193, 6350, 8264, 6819, 6921, 6934, 7332, 7341, 7362, 7733, 7828, 7858, 7933, 7934, 8011, 8155, 8290, 6383, 5182, 5219, 5296, 5398, 8157, 6760, 7723, 10108, 8365, 8367, 10504, 10657, 10501, 4274, 8046, 8048, 5892, 7818, 4773, 5585, 8113, 5165, 4402, 6468, 6319, 5313, 7511, 4952, 5277, 5457, 4116, 5403, 5078, 5664, 5666, 6327, 6733, 6994, 6136, 6206, 6172, 7604, 7602, 7353, 6273, 6993, 4045, 10221, 10261, and 8540.

Distribution Pattern:

OUS: Japan, Belgium, Denmark, France, Great Britain, and Spain. No U.S. consignees.

Voluntary or Mandated:

Voluntary: Firm initiated