Boston Scientific Corporation: Medical Device Recall in 2020 - (Recall #: Z-2682-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Ablation Galil Technology, IcePearl 2.1 CX 90 Cyroablation Needles, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures.

Product Classification:

Class II

Date Initiated: June 4, 2020
Date Posted: July 29, 2020
Recall Number: Z-2682-2020
Event ID: 85815
Reason for Recall:

The needle surface deteriorated due to unexpected electrolysis during procedure.

Status: Terminated
Product Quantity: 205 devices
Code Information:

UPN: FPRPR3617, GTIN: 07290104831072 , Batch: A0904, A1085, A1177, A1178, A6681, A6682, A6962, A6963, A6977, A7113, U0084

Distribution Pattern:

Nationwide Distribution to states of: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, KS, LA, MA, MD, ME, MN, MO, MS, NC, ND, NE, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI WV, and WY and the country of: Canada.

Voluntary or Mandated:

Voluntary: Firm initiated