Boston Scientific Corporation: Medical Device Recall in 2023 - (Recall #: Z-2228-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Flexiva Pulse 242 Single-Use Laser Fibers-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Single-pack UPN : M006L8405910
Product Classification:
Class II
Date Initiated: May 25, 2023
Date Posted: August 2, 2023
Recall Number: Z-2228-2023
Event ID: 92570
Reason for Recall:
Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched
Status: Ongoing
Product Quantity: 66 units
Code Information:
GTIN: 08714729978572 Lot Number: 30554442 Exp. Date: 16-Nov-2025
Distribution Pattern:
Worldwide distribution - US Nationwide and the country of Canada.
Voluntary or Mandated:
Voluntary: Firm initiated