Boston Scientific Corporation: Medical Device Recall in 2023 - (Recall #: Z-2232-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy (Box of 5)UPN: M006L8406911
Product Classification:
Class II
Date Initiated: May 25, 2023
Date Posted: August 2, 2023
Recall Number: Z-2232-2023
Event ID: 92570
Reason for Recall:
Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched
Status: Ongoing
Product Quantity: 205 units
Code Information:
GTIN: 08714729978725 Lot Numbers: 30521480, 30521485 ,30521486 Exp. Date: 20-OCT-2024 to 10-Nov-2025
Distribution Pattern:
Worldwide distribution - US Nationwide and the country of Canada.
Voluntary or Mandated:
Voluntary: Firm initiated