Boston Scientific Corporation: Medical Device Recall in 2024 - (Recall #: Z-1153-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.
Product Classification:
Class II
Date Initiated: December 19, 2023
Date Posted: February 28, 2024
Recall Number: Z-1153-2024
Event ID: 93794
Reason for Recall:
Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.
Status: Ongoing
Product Quantity: 0 US; 74 OUS
Code Information:
UPN M00509500, GTIN (UDI-DI) 08714729902713, Lot 31073152; UPN M00509510, GTIN (UDI-DI) 08714729902720, Lots 31221751, 31761882, 32409573.
Distribution Pattern:
International distribution to the country of Japan only.
Voluntary or Mandated:
Voluntary: Firm initiated