Boston Scientific Corporation: Medical Device Recall in 2024 - (Recall #: Z-1515-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
EndoVive 20Fr Push Safety PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00566471
Class II
Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration
GTIN: 08714729802631 Lot Numbers: 32295181 , 32295182 , 32306054 , 32306055 , 32316800 , 32323673 , 32342013 ,32367049 , 32367050 , 32374782 , 32472868 , 32486230 , 32486231 , 32486232 , 32486233 , 32506056 , 32515039 , 32523243 , 32523244 , 32531793 , 32531794 , 32531795 , 33113026 , 33113027 , 33113028 , 33113029 , 33170338
Worldwide distribution - US Nationwide and the country of Japan.
Voluntary: Firm initiated