Boston Scientific Corporation: Medical Device Recall in 2025 - (Recall #: Z-2326-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296

Product Classification:

Class I

Date Initiated: July 24, 2025
Date Posted: August 27, 2025
Recall Number: Z-2326-2025
Event ID: 97318
Reason for Recall:

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Status: Ongoing
Product Quantity: 247,702 US; 344,670 OUS
Code Information:

Model 0282 UDI-DI/GTIN 00802526431371; Model 0283 UDI-DI/GTIN 00802526431388 00802526504181; Model 0285 UDI-DI/GTIN 00802526431401; Model 0286 UDI-DI/GTIN 00802526431418; Model 0292 UDI-DI/GTIN 00802526431432 00802526474668 00802526480096 00802526485763 00802526504396 00802526531248 00802526541742 00802526552441 00802526611186; Model 0293 UDI-DI/GTIN 00802526431449 00802526531255 00802526541827 00802526552526 00802526554490 00802526611179; Model 0295 UDI-DI/GTIN 00802526431463 00802526474699 00802526531262 00802526541902 00802526554506 00802526611131; Model 0296 UDI-DI/GTIN 00802526431470 00802526474705 00802526480164 00802526480188 00802526531279 00802526541988 00802526552038

Distribution Pattern:

Worldwide Distribution. US nationwide. International distribution worldwide.

Voluntary or Mandated:

Voluntary: Firm initiated