Boston Scientific Neuromodulation Corporation: Medical Device Recall in 2025 - (Recall #: Z-2558-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Class II
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
Voluntary: Firm initiated