Breas Medical, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0943-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
Product Classification:
Class II
Date Initiated: March 21, 2022
Date Posted: April 20, 2022
Recall Number: Z-0943-2022
Event ID: 89776
Reason for Recall:
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
Status: Terminated
Product Quantity: 707 units
Code Information:
Trolleys delivered between June 11th, 2020, and December 3rd, 2021. UDI: 07321820073849
Distribution Pattern:
US Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated