Bridge to Life Ltd: Medical Device Recall in 2025 - (Recall #: Z-2157-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250
Class II
The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.
UDI 10860001991205-240206-260206-240372708 (case), 00860001991208-240206-260206-240372708 (bag), Lot Number 240372708; UDI 10860001991205-240212-260212-240402836 (case), 00860001991208-240212-260212-240402836 (bag), Lot Number 240402836; UDI 10860001991205-240214-260214-240402837 (case), 00860001991208-240214-260214-240402837 (bag), Lot Number 240402837.
United States - New York, Maryland, Alabama, North Carolina, South Carolina, Ohio, Florida, Wisconsin, Utah, Tennessee, New Jersey, Georgia, Nebraska, Arizona, Minnesota, California, and Canada
Voluntary: Firm initiated