Burlington Medical, LLC: Medical Device Recall in 2019 - (Recall #: Z-2117-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Peak Vest, Single Item, Product Number PVest, Part Numbers REPVEST*M, SEPVESTE, SEPVEST*F Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Product Classification:

Class II

Date Initiated: June 7, 2019
Date Posted: August 7, 2019
Recall Number: Z-2117-2019
Event ID: 83144
Reason for Recall:

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Status: Terminated
Product Quantity: 7
Code Information:

Serial Numbers: 418892 381887 381902 418811 418952 387916 429678

Distribution Pattern:

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated