California Medical Laboratories Inc: Medical Device Recall in 2013 - (Recall #: Z-0650-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Rigid Tip Suction Wand Non sterile bulk devices (S112468N, S120196N, and S120197N) were sold to Terumo UK to be included in a kit under Terumo as manufacturer's name. The Suction Wands are indicated for use to remove excess fluid from the surgical field.
Class II
California Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the potential to come loose from the wand and fall into the patient's chest cavity.
Catalog No: SU-12202NS Lot No: S112468N, S120196N, S120197N
Worldwide Distribution - USA including AZ, FL, CA, LA, CO, and MS. Internationally to Canada, Israel, Greece, Hong Kong, India, Ireland, Jordan, Lebanon, Malaysia, Netherlands, Russia, Saudi Arabia, South Africa and Spain.
Voluntary: Firm initiated