Cambridge Sensors Limited: Medical Device Recall in 2019 - (Recall #: Z-2357-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Microdot Xtra blood glucose test strips
Product Classification:
Class II
Date Initiated: July 26, 2018
Date Posted: August 28, 2019
Recall Number: Z-2357-2019
Event ID: 82867
Reason for Recall:
The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Status: Ongoing
Product Quantity: 17,784 (USA)
Code Information:
LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.
Distribution Pattern:
Illinois
Voluntary or Mandated:
Voluntary: Firm initiated