Canon Medical System, USA, INC.: Medical Device Recall in 2019 - (Recall #: Z-0360-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Product Classification:

Class II

Date Initiated: February 8, 2019
Date Posted: November 20, 2019
Recall Number: Z-0360-2020
Event ID: 84135
Reason for Recall:

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Status: Terminated
Product Quantity: 9 systems
Code Information:

Model CAT-870B D053991 Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.

Voluntary or Mandated:

Voluntary: Firm initiated