Canon Medical System, USA, INC.: Medical Device Recall in 2020 - (Recall #: Z-0532-2021)
See the recall detail below. You can also see other recalls from the same firm in 2020.
CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Class II
There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.
Affected Serial Numbers: 1 AFA1662001, 2 AFB1872005, 3 AFB1832003, 4 AFB1872004, 5 AFB1892006, 6 BFA18Y7001, 7 AEA17X2001, 8 AFA1692002, 9 AEB1812002, 10 BFA1932001, 11 BFB1952003, 12 BFB1942002, 13 BFB1972004, 14 BFB1972005, 15 AFB1892007, 16 BEC2032001, 17 BFC2062007, 18 BFC2062008, 19 BHA1992001, 20 BFC2012006
US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.
Voluntary: Firm initiated