Canon Medical System, USA, INC.: Medical Device Recall in 2020 - (Recall #: Z-0549-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.

Product Classification:

Class II

Date Initiated: November 9, 2020
Date Posted: December 16, 2020
Recall Number: Z-0549-2021
Event ID: 86864
Reason for Recall:

A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.

Status: Terminated
Product Quantity: 60 units
Code Information:

Serial Numbers: 8LA2052063, 8LA19Z2034 , 8LA19X2004, 8LA19Z2027, 8LA19Z2029, 8LA2052062, 8LA2042055, 8LA19Z2032, 8OA2052059, 8LA2042051, 8LA2032044, 8LA19X2002, 1LB1922175, 8LA2042050, 8LA19X2007, 8LA2012037, 8LA19Y2021, 8LA19Y2014, 8LA2042056, 8LA19Z2023, 8LA2052067, 8LA19Z2031, 8LA19Y2019, 8LA19Z2030, 8OA19Y2013, 8LA19Z2028, 8LA19Y2012, 8LA19Y2020, 8LA2032042, 8LA2052066, 8LA19X2003, 8LA19Y2022, 8LA19X2008, 8LA2042047, 8LA19Z2024, 8LA2032046, 8LA19Z2033, 8OA2042053, 8LA2012036, 8LA2042054, 8LA19Z2025, 8LA19Z2026, 8LA2022041, 8LA19X2010, 8LA2032043, 8OA19X2001, 8OA2052061, 8OA2042049, 8LA19Y2016, 8LA2022039, 8LA19Y2015, 8LA2012035, 8LA2032045, 8LA19Y2018, 8LA19X2005, 8LA19X2006, 8LA19X2009, 8LA19Y2011, 8LA19Y2017, 8LA2012038, 8LA2052060, 8LA2052058 , 8LA2052064, 8LA2052065, 8LA2052068, 8LA2022040, 8OA2042052, 8LA2042048, 8LA2052057

Distribution Pattern:

US Nationwide distribution. OUS: Unknown

Voluntary or Mandated:

Voluntary: Firm initiated