Canon Medical System, USA, INC.: Medical Device Recall in 2020 - (Recall #: Z-2326-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Infinix-8000C with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Class II
Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Serial Numbers: B8B1972001, A9A1592001, A3A14Z2001, AEB1812002, A1B1552001, A9C18Y2007, A3E1922002, B5B1952001, A9C1822006, A9B1622002, A9B1722003, A5C1622001, B3B19Y2001, A1D1772003, B1B19X2001, A5C16Y2002, A9C1822005, A1E1862008, A1D1752002, A1E1832004, A1E1852005, A1E1852006, A5E1862003, AEA17X2001
US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.
Voluntary: Firm initiated