Canon Medical System, USA, INC.: Medical Device Recall in 2021 - (Recall #: Z-0142-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A
Product Classification:
Class II
Date Initiated: November 11, 2020
Date Posted: October 27, 2021
Recall Number: Z-0142-2022
Event ID: 88691
Reason for Recall:
A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.
Status: Terminated
Product Quantity: 3 units
Code Information:
Affected serial numbers: 3DA1982001, 3DA2072002, 3DA2072003
Distribution Pattern:
US Nationwide distribution in the states of NV, AR MN.
Voluntary or Mandated:
Voluntary: Firm initiated