Canon Medical System, USA, INC.: Medical Device Recall in 2023 - (Recall #: Z-2392-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE
Class II
The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.
UDI-DI: 04987670100192. INFX-8000C, Serial Numbers: BHA1992001, BHB2132003, SWC2242002, AFB1832003, SUC2262008, SUC2292010, BFB1972004, SUB2172002, SWC2262003
US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA
Voluntary: Firm initiated