Canon Medical System, USA, INC.: Medical Device Recall in 2024 - (Recall #: Z-1249-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
MRI system: Vantage Elan, Model: MRT-2020
Class II
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
UDI-DI: 04987670101632. Serial Number: S1F2052753
Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.
Voluntary: Firm initiated