Canon Medical System, USA, INC.: Medical Device Recall in 2024 - (Recall #: Z-1250-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Vantage Titan, Model: MRT-1504

Product Classification:

Class II

Date Initiated: January 29, 2024
Date Posted: March 13, 2024
Recall Number: Z-1250-2024
Event ID: 93874
Reason for Recall:

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Status: Ongoing
Product Quantity: 40
Code Information:

UDI-DI: 04987670102400. Serial Numbers: U4B15Y2102, U5B15Z2001, U4B1492080, U4B1582100, U4B1582099, U4B1522091, U4B1512089, U4B1432067, U4B1522093, U4B1452073, U4B1442069, U4B1442068, U4B1642106, U4B1432066, U4B15X2101, U4B14Y2083, U4B1452071, U4B1512087, U4B1652107, U4B1552097, U4B1482078, U4B1512090, U4B1552098, U4B1442070, U4B14Y2082, U4B1482076, U4B14Z2085, U4B14Z2086, U4B1512088, U4B1612103, U4B1472075, U4B1422065, U4B1462074, U4B14Y2084, U4B1632105, U4B1452072, U4B1492081, U4B1482079, U4B1612104, U4B1482077

Distribution Pattern:

Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.

Voluntary or Mandated:

Voluntary: Firm initiated