Canon Medical System, USA, INC.: Medical Device Recall in 2025 - (Recall #: Z-0327-2026)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Product Classification:
Class II
Date Initiated: October 16, 2025
Date Posted: November 5, 2025
Recall Number: Z-0327-2026
Event ID: 97735
Reason for Recall:
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Status: Ongoing
Product Quantity: 2 units
Code Information:
CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.
Distribution Pattern:
US distribution to Ohio and Nevada
Voluntary or Mandated:
Voluntary: Firm initiated