Cardiac Assist, Inc: Medical Device Recall in 2020 - (Recall #: Z-1106-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
TandemLife ECLS Procedure Kit - V24/A17, Catalog Number 5740-2417 - Product Usage: The function of the 24Fr ProtekSolo Venous Cannula is to provide a conduit for blood to travel from the patient s venous system when connected to accessory extracorporeal equipment. The Venous Cannula is intended to cannulate vessels, perfuse vessels or organs and/or connect with accessory extracorporeal circulatory support equipment. The introducer is intended to facilitate proper insertion and placement of the cannula within the vessel for extracorporeal circulatory support. These devices are to be used by a trained physician only.
Class II
The kit contained a 29Fr ProtekDuo cannula instead of a 24Fr Protek Solo Venous cannula.
Serial Number/UDI 141445/ (01)00814112020517(17)200825(11)190423(21)00141445, 141446/ (01)00814112020517(17)200825(11)190423(21)00141446, 141447/ (01)00814112020517(17)200825(11)190423(21)00141447, 141448/ (01)00814112020517(17)200825(11)190423(21)00141448, 141449/ (01)00814112020517(17)200825(11)190423(21)00141449, 141450/ (01)00814112020517(17)200825(11)190423(21)00141450, 141451/ (01)00814112020517(17)200825(11)190429(21)00141451, 141452/ (01)00814112020517(17)200825(11)190429(21)00141452, 141453/ (01)00814112020517(17)200825(11)190429(21)00141453, 141454/ (01)00814112020517(17)200825(11)190429(21)00141454, 141457/ (01)00814112020517(17)200825(11)190429(21)00141457 Expiration Date for all products - August 25, 2020
US Nationwide distribution in the states of AZ, AL, ME, LA, PA.
Voluntary: Firm initiated