Cardiac Science Corporation: Medical Device Recall in 2016 - (Recall #: Z-0829-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

9131-001 Defibrillation Electrodes, Lot No. 141125-02 Used in Powerheart G3 9300A, 9300E, 9300P, 9390A, and 9390E AEDs Cardiac Science 9131 Defibrillation Electrodes are single use and intended to be used in conjunction with Cardiac Science automatic external defibrillators (AED) to monitor and deliver defibrillation energy to the patient. The electrodes are intended for short term use (<8 hours) and must be used before the expiration date listed on the packaging. The AED electrodes are used for emergency treatment of cardiac arrest patients over 8 years of age or greater than 55 pounds. The user assesses the patient s condition and confirms that the patient is unconscious, pulseless and is not breathing prior to applying the electrodes to the skin.

Product Classification:

Class II

Date Initiated: January 14, 2016
Date Posted: March 2, 2016
Recall Number: Z-0829-2016
Event ID: 72824
Reason for Recall:

Defibrillation electrodes may increase electrical impedance over time. If impedance becomes too high, the AED will fail the electrode self-test and it will not be Rescue Ready, the AED STATUS INDICATOR will be RED and the AED will beep. This is normal behavior when the impedance limit is exceeded. The high impedance electrodes require replacement.

Status: Terminated
Product Quantity: 10,943
Code Information:

Lot No. 141125-02

Distribution Pattern:

Worldwide Distribution - US all states including PR except ND & RI. International Distribution to Argentina, Australia, Belgium, Bermuda, Canada, Chile, Colombia, Ecuador, France, Germany, Hong Kong, Iceland, Ireland , Israel, Lebanon, Malaysia, Malta, Norway, Oman, Philippines, Poland, Qatar, Romania, Russian Federation, Singapore, Slovenia, Sweden, Turkey, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated